Memo /
ID06PressSoftware
This tutorial is part of the ID06 memo
Starting a compression
All of this is done in the Voggenreiter controller
- go into
Profileand set a ramp, for instance- 10 mn to reach 10 bars,
- 60 mn for 30 bars,
- and finally 2000 mn at 30 bars as final step
- go into
Settings- check
Data storageand that you have a.datlog file - make sure that the ckeckbutton
saveis selected
- check
- go into
Controls,Active distance controlshould beOn,Hold followshould beHold position,Hold position ramp timeshould be in agreement with compression times you chose inpofile, 4200 sec for 70 mn, for instance,End valueis the position that differentials rams should go to, to be determined with Jeremy.
- Go into
Profile- Menu
PLC controller Download profile to controller- Menu
PLC controller Run continue profile
- Menu
- Go into
Controls, activateRamp On/Off
The press should be going up in pressure...
Start a deformation
If you are at pressure and want to start a deformation....
- Open the application:
press monitor - Open the
Controlstab- Look at differential rams in
Controltab - choose
end valueandtime - example: remove 0.350 mm in 2h for 1 to 3.10-5 s-1
- Look at differential rams in
- Open the
ProfiletabSuspend profileinPLC Controllermenu- remove about 2% of pressure on main ram over 2 hours, plus a long plateau afterward
Download profileinPLC ControllermenuRun profileinPLC Controllermenu
- Back to differential rams in
ControltabActive distance controlshould beOn,Hold followshould beHold position,- Select
Rampto start deformation
You are good to go.